Application of a novel Revolver apparatus for a biorelevant dissolution test of ropinirole sustained release tablets

 

Macedonian pharmaceutical bulletin, 69 (Suppl 1) 241 – 242 (2023)
Online ISSN 1857 – 8969
DOI: 10.33320/maced.pharm.bull.2023.69.03.118
Short communication

Application of a novel Revolver apparatus for a biorelevant dissolution test of
ropinirole sustained release tablets

Bartłomiej Milanowski, Mateusz Kuprianowicz
GENERICA R&D Lab, Regionalne Centrum Zdrowia Sp. z o.o., 3 Na Kepie St., 64-360 Zbaszyn, Poland

 

 

Introduction
Biorelevant dissolution tests of oral solid dosage forms open the gate to valid in vitro-in vivo predictions (IVIVP). A recently developed Revolver apparatus (Fig. 1) allows mimicking the acceptor fluid composition and flow as well as pH changes occurring in the human fasted intestine. This work presents and showcases the first application of the Revolver in dissolution testing of SR tablets with ropinirole hydrochloride (RH).

Fig. 1. The Revolver apparatus

Мaterials and methods
RH secondary pharmaceutical standard (99.5 % pure) was purchased from Sigma-Aldrich (Poland). Polpix SR 2 mg tablets were purchased from a local pharmacy. All other chemicals were at the analytical grade.
Drug dissolution profiles from Polpix SR 2 mg tablets were obtained according to modified USP Ropinirole Extended-Release Tablets test 1 protocol by Revolver apparatus (Physiolution Polska Sp. z o.o., Poland). Tablets in JP sinkers were placed in a vessel of Revolver apparatus (Test Scenario A, Fig. 2). In the Test Scenario B, the tablets were positioned in the 3D printed flow through cells (Fig. 1 and 2). The paddle speed of Revolver was fixed at 100 rpm (Scenario A) and the flow rate of peristaltic pump was set at 8 ml/min (Scenario B). The tests were conducted in 900 mL of citric acid buffer pH=4.0 or 4 mM modified Hanks’ buffer at 37 ± 0.5 °C for 24 h. The pH profiles in modified Hanks’ buffer were simulated and controlled by integrated, 3-channel pHysio-grad® v. 2.2 (Physiolution Polska Sp. z o.o., Poland) which enables a dynamic adjustment and change of the media pH by introducing CO2 or an inert gas into the dissolution medium (Fig. 2). Samples (1 mL) were withdrawn at 1, 2, 4, 6, 8, 12, 16, 20 and 24 h and filtered through 1 μm UHMW PE cannula filters (Dissolution Accessories, The Netherlands).

Fig. 2. Scheme of ropinirole dissolution test scenarios of Polpix SR 2 mg tablets in the Revolver apparatus.

 

The amount of released RH was determined using a validated RP-HPLC-UV method using a Kinetex® XB-C18 column (2.6 μm, 100 Å, 3 x 50 mm, Phenomenex,
242 Pharmaceutical analysis and bioanalysis United States). The mobile phase was composed of acetate buffer pH=2.5, ACN and MeOH (800:140:60, v/v/v). Injections of 8 μL were made and HPLC oven and PDA detector temperatures were 35 and 40 °C, respectively.

Results and discussion
RH dissolution profiles from Polpix SR 2 mg tablets in the conditions of Test Scenario A and B are shown in the Fig. 3A and 3B, respectively. The fastest dissolution kinetics of RH was demonstrated when Polpix SR tablets were placed in JP sinkers directly in Revolver vessel with a paddle (Fig. 3A). Slightly slower and incomplete dissolution of RH was observed when tablets were positioned in the 3D printed flow through cells (Fig. 3B). We do not observed the influence of media type on RH dissolution in both scenarios (f2 > 50

Fig. 3. The RH dissolution profiles from Polpix SR 2 mg tablets in the Revolver apparatus under: A) Test Scenario A, and B) Test Scenario B. The f2 values were calculated to compare the release profiles in various media.

Conclusion
The innovative Revolver device provides various hydrodynamic conditions and enables a dynamic adjustment and change of the bicarbonate buffers media pH to evaluate the dissolution efficiency of a sustained release dosage form. It can reflect the hydrodynamic conditions of the USP dissolution apparatus 2 (paddle) or 4 (flow-through cell). Each station is an independent unit, which can work flexibly with different dissolution test conditions and parameters. Moreover, it is cheap and thus can be a useful tool in pharmaceutical R&D labs for development of new products and to compare products properties under bio-relevant test conditions.

Acknowledgements: The authors are grateful for the financial support received within the Polish National Centre for Research and Development grant number POIR.04.01.04‐00‐0142/17.

References
Garbacz, G., Kołodziej, B., Koziolek, M., Weitschies. W., Klein, S., 2014. A dynamic system for the simulation of fasting luminal pH-gradients using hydrogen carbonate buffers for dissolution testing of ionisable compounds. Eur J Pharm Sci. 51, 224-31. doi: 10.1016/j.ejps.2013.09.020.
Shuklinova, O., Dorożyński, P., Kulinowski, P., Polak, S., 2022. Quality Control Dissolution Data Is Biopredictive for a Modified Release Ropinirole Formulation: Virtual Experiment with the Use of Re-Developed and Verified PBPK Model. Pharmaceutics 14, 1514. https://doi.org/10.3390/pharmaceutics14071514

 

DOWNLOAD PDF:
Application of a novel Revolver apparatus for a biorelevant dissolution test of ropinirole sustained release tablets